Edman degradation
Edman degradation, developed by Pehr Edman, is a method of sequencing amino acids in a peptide.[1] In this method, the amino-terminal residue is labeled and cleaved from the peptide without disrupting the peptide bonds between other amino acid residues.
Mechanism
Phenyl isothiocyanate is reacted with an uncharged N-terminal amino group, under mildly alkaline conditions, to form a cyclical phenylthiocarbamoyl derivative. Then, under acidic conditions, this derivative of the terminal amino acid is cleaved as a thiazolinone derivative. The thiazolinone amino acid is then selectively extracted into an organic solvent and treated with acid to form the more stable phenylthiohydantoin (PTH)- amino acid derivative that can be identified by using chromatography or electrophoresis. This procedure can then be repeated again to identify the next amino acid. A major drawback to this technique is that the peptides being sequenced in this manner cannot have more than 50 to 60 residues (and in practice, under 30). The peptide length is limited due to the cyclical derivatization not always going to completion. The derivatization problem can be resolved by cleaving large peptides into smaller peptides before proceeding with the reaction. It is able to accurately sequence up to 30 amino acids with modern machines capable of over 99% efficiency per amino acid. An advantage of the Edman degradation is that it only uses 10 - 100 pico-moles of peptide for the sequencing process. The Edman degradation reaction was automated in 1967 by Edman and Beggs to speed up the process[2] and 100 automated devices were in use worldwide by 1973.[3]
Limitations
Because the Edman degradation proceeds from the N-terminus of the protein, it will not work if the N-terminal amino acid has been chemically modified or if it is concealed within the body of the protein. It also requires the use of either guesswork or a separate procedure to determine the positions of disulfide bridges, and peptide concentrations of 1 picomolar or above for discernible results.
Coupled analysis
Following 2D SDS PAGE the proteins can be transferred to a polyvinylidene difluoride (PVDF) blotting membrane for further analysis. Edman degradations can be performed directly from a PVDF membrane. N-terminal residue sequencing resulting in five to ten amino acid may be sufficient to identify a Protein of Interest (POI).
See also
References
- ↑ Edman, P.; Högfeldt, Erik; Sillén, Lars Gunnar; Kinell, Per-Olof (1950). "Method for determination of the amino acid sequence in peptides". Acta Chem. Scand. 4: 283–293. doi:10.3891/acta.chem.scand.04-0283.
- ↑ Edman P, Begg G (March 1967). "A protein sequenator". Eur J Biochem. 1 (1): 80–91. PMID 6059350.
- ↑ Niall HD (1973). "Automated Edman degradation: the protein sequenator". Meth. Enzymol. Methods in Enzymology. 27: 942–1010. doi:10.1016/S0076-6879(73)27039-8. ISBN 978-0-12-181890-6. PMID 4773306.