Indirect calorimetry
Indirect calorimetry calculates heat that living organisms produce by measuring either their production of carbon dioxide and nitrogen waste (frequently ammonia in aquatic organisms, or urea in terrestrial ones), or from their consumption of oxygen. Indirect calorimetry is the method by which the type and rate of substrate utilization, and energy metabolism are estimated in vivo starting from gas exchange measurements (carbon dioxide production and oxygen consumption during rest and steady-state exercise). This technique provides unique information, is noninvasive, and can be advantageously combined with other experimental methods to investigate numerous aspects of nutrient assimilation, thermogenesis, the energetics of physical exercise, and the pathogenesis of metabolic diseases.[1]
Scientific Background
Indirect calorimetry measures O2 consumption and CO2 production. On the assumption that all the oxygen is used to oxidize degradable fuels and all the CO, thereby evolved is recovered, it is possible to calculate the total amount of energy produced. It should be clear that “energy production” means conversion of the chemical free-energy of nutrients into the chemical energy of ATP plus loss of some energy during the oxidation process.[1] Respiratory indirect calorimetry, or only indirect calorimetry (IC) as it is often known by most authors, is a noninvasive and highly accurate method of metabolic rate which has an error rate lower than 1%.[2] It has high reproducibility and has been considered a gold standard method.[3] This method allows estimating BEE and REE, and also allows for identification of energy substrates that are being predominantly metabolized by the body at a specific moment. It is based on the indirect measure of the heat produced by oxidation of macronutrients, which is estimated by monitoring oxygen consumption (O2) and carbon dioxide production (CO2) for a certain period of time.[4] The calorimeter has a gas collector that adapts to the subject and a system that measures the volume and concentrations of O2 and CO2 minute by minute. Through a unidirectional valve, the calorimeter collects and quantifies the volume and concentration of O2 inspired and CO2 expired by the subject. After a volume is met, Resting Energy Expenditure is calculated by the Weir formula and results are displayed in software attached to the system.[4]
History
Antoine Lavoisier noted in 1780 that heat production can be predicted from oxygen consumption, using multiple regression. The Dynamic Energy Budget theory explains why this procedure is correct. Indirect Calorimetry, as we know it, was developed at the turn of the century’s as an application of thermodynamics to animal life.[5] Although the development of indirect calorimetry dates back over 200 years, its greatest use has been in the last two decades with the development of total Parenteral nutrition, interdisciplinary nutrition support teams, and the production of portable, reliable, relatively inexpensive calorimeters.[6]
Collection Methods
Four different gas collection and measurement techniques can be used to perform this test:
- Douglas Bag: expired respiratory gases are collected on an inflatable airtight bag.[7] After completion of any test using Douglas Bags, gas collected must be analysed for volume and composition.
- Canopy (dilution): The dilution technique is considered the gold standard technology for Resting Energy Expenditure measurement in clinical nutrition.[3] The test lasts just few minutes and consists of making a patient lie down relaxed on a bed or on a comfortable couch, with the head under a transparent hood connected to a pump, which applies an adjustable ventilation through it. Exhaled gas dilutes with the fresh air ventilated under the hood and a sample of this mixture is conveyed to the analysers, through a capillary tube and analysed. Ambient and diluted fractions of O2 and CO2 are measured for a known ventilation rate, and O2 consumption and CO2 production are determined and converted into Resting Energy Expenditure.[8]
- Face mask (breath by breath): Indirect calorimetry tests are also often performed with a face mask, which is used to convey exhaled and inhaled gas through a turbine flowmeter able to measure the patient’s breath by breath minute ventilation, at the same time a sample of gas is conveyed to the analyser and VO2 and VCO2 are measured and converted in energy expenditure.
- Interface with a Ventilator (Intensive Care settings): In case the patient is mechanically ventilated, an indirect calorimeter can still measure breath by breath inhaled/exhaled O2 and CO2 if interfaced with the ventilator through the endotracheal tube.
Applications
Indirect calorimetry provides at least two pieces of information: a measure of energy expenditure or 24-hour caloric requirements as reflected by the Resting Energy Expenditure (REE) and a measure of substrate utilization as reflected in the Respiratory Quotient (RQ). Knowledge of the many factors that affect these values has led to a much broader range of applications. Studies of indirect calorimetry over the past 20 years have led to the characterization of the hypermetabolic stress response to injury and the design of nutritional regimens whose substrates are most efficiently assimilated in different disease processes and organ failure states. Indirect calorimetry has influenced everyday practices of medical and surgical care, such as the warming of burn unit and surgical suites and the weaning of patients from ventilators.[6]
References
- 1 2 Ferrannini E."The theoretical bases of indirect calorimetry: a review." Metabolism. 1988 Mar;37(3):287-301.
- ↑ Marson F, et al. "Correlation between oxygen consumption calculated using Fick's method and measured with indirect calorimetry in critically ill patients." Arq Bras Cardiol. 2004 Jan;82(1):77-81, 72-6. Epub 2004 Feb 12.
- 1 2 Haugen HA, et al. "Indirect calorimetry: a practical guide for clinicians." Nutr Clin Pract. 2007 Aug;22(4):377-88.
- 1 2 Pinheiro Volp AC, et al. "Energy expenditure: components and evaluation methods." Nutr Hosp. 2011 May-Jun;26(3):430-40. doi: 10.1590/S0212-16112011000300002.
- ↑ Atwater WO, et al. "Description of neo respiration calorimeter and experiments on the conservation of energy in the human body." US Department Agriculture, Off Exp Sta Bull 63, 1899
- 1 2 McClave SA, et al. "Use of indirect calorimetry in clinical nutrition." Nutr Clin Pract. 1992 Oct;7(5):207-21.
- ↑ Douglas, C. Gordon (18 March 1911). "A method for determining the total respiratory exchange in man". Proceedings of the Physiological Society. (Douglas Bag)
- ↑ Academy of Nutrition and Dietetics "Measuring RMR with Indirect Calorimetry (IC)." Nutr Clin Pract. 2007 Aug;22(4):377-88.
See also
- "Measuring RMR with Indirect Calorimetry (IC)" Academy of Nutrition and Dietetics - Evidence Analysis Library
- American Journal of Clinical Nutrition